Product Description
Condumax II is a fully self-contained hydrocarbon dew point analyzer, using the proven “Dark Spot” hydrocarbon dew point measurement principle from the highly successful Condumax instrument. Combining a hydrocarbon dew point cell and water dew point measurement in the same compact and lightweight EExd enclosure, Condumax II requires nothing more than mains power to operate.
Product Features
- Fully automatic on-line analysis
- Objective, highly repeatable
- 0.5°C hydrocarbon dew-point accuracy
- Fundamental chilled mirror principle
- Patented detection technique
- Self-cleaning
- No purge or cooling gas needed
- Optional water dew-point analysis
- Modbus RTU
- IECEx, ATEX, CCSAUS and GOST Ex certifications
Technical Specification
-
Hydrocarbon Dew-Point Measurement Measuring Technique Dark SpotTM fixed sample analysis.
Direct photo-detection of hydrocarbon condensate at hydrocarbon dew-point temperatureSensor Cooling Automatic via 3-stage Peltier effect electronic cooler under adaptive control Maximum Range Up to Δ -55K measurement depression from Main Unit operating temperature Accuracy ±0.5°C hydrocarbon dew point (single and multiple condensible component analysis) Sample Flow 0.03m³/hr (0.5 Nl/min) – alarm standard Measurement Frequency 6 cycles/hour (recommended)
12 cycles/hour (maximum)Operating Pressure Max 100 barg (1450 psig) Water Dew-Point Measurement (Optional) Measuring Technique Michell Ceramic Moisture Sensor Units – Moisture content °C and °F water dew point; lb/MMscf; mg/m3, ppmV Resolution 0.1°C, 0.1°F Range Calibrated from -100 to +20°C dew point Accuracy ±1°C from -59 to +20°C dew point
±2°C from -100 to -60°C dew pointSample Flow 0.06 to 0.3 m3/hr (1 to 5 Nl/min) alarm standard Temperature Coefficient Algorithm compensation (-20 to +50°C) Operating Pressure Max 138 barg (2000 psig) Pressure Measurement
Hydrocarbon Dew-Point Analysis Pressure
Optional Water Dew-Point Process PressureUnits MPa, barg, psig Resolution 0.1 MPa and barg, 1 psig Range HC dew point: 0 to 100 barg
W dew point: 0 to 200 bargAccuracy ±0.25% FS Hydrocarbon Dew-Point Analyzer Resolution Hydrocarbon and water dew point: 0.1°C (0.18°F) Sample Gas Supply Natural gas up to 100 barg (1450 psig), pressure regulated in sampling system Enclosure EExd cast enclosure with removable glass window viewing port. Internally heated for condensation protection Sample Gas Connections 1/8” NPT (ATEX / IECEx) or 1/4” NPT (CSA) female ports for both hydrocarbon and water dew-point channels; Sampling system: 6mm OD or 1/4” OD Operating Environment Indoors/Outdoors -20 to +50°C (-4 to +122°F); Max 95% RH Power Supply 90 to 260 V AC, 50/60 Hz, 125 W Main Unit
300 W c/w Sampling System (indoor)
400 W c/w Sampling system (outdoor)Weight Main Unit 22.5kg (49.6lbs)
~60kg (132.3lbs) c/w Sampling system (indoor)
~75kg (165.3lbs) c/w Sampling system (outdoor)Integrated Display/Keyboard Touch screen with vacuum fluorescent display Outputs Modbus RTU, RS485 @ 9600 baud rate
Two 4-20 mA linear (non-isolated) outputs, user-configurable for any combination of dew-point or pressure parametersAlarms Process and analyzer status via software register and display annotation
Integrated low flow alarms for each sample flow
Analyzer status fault flag 23 mA on mA Output 1Certification Hazardous Area Certification ATEX: II 2G Ex d IIB + H2 Gb
IECEx: Ex d IIB + H2 Gb
Tamb -40°C to +45°C T4
Tamb -40°C to +60°C T3
cCSAus: Class I, Division 1,
Group B, C, & D T4
*TC TR: 1Ex d IIB+H2 T4, T3 GbPattern Approval GOST-R, GOST-K, GOST-T
Any Question at
+92 51 4862395
Office # 06, Rose-1 Plaza
I-8 Markaz, Islamabad
Mon-Sat
9 am – 5 pm
Send your mail at
info@tetracon.com.pk