Product Description
The Michell HG10 Humidity Calibration System is a highly flexible computer-controlled automatic calibration system for humidity sensors. The HG10 is capable of repeatable generation of temperature and relative humidity set points over the range 1 to 95% RH
(-50 to +50°Cdp) at temperatures of +20 to +50°C with excellent stability. The supplied chilled mirror reference instrument provides traceability directly to national standards, and makes the system suitable for use in high-level calibration laboratories.
The supplied PC software allows the creation of automatic calibration programs, for evaluating the performance of humidity sensors over a range of operating conditions. The measured values from the chilled mirror reference are used in a closed control loop to enable repeatable set point generation, time and time again.
Product Features
- Automatic calibration system with reference hygrometer
- Wide range of humidity generation 1 to 95% RH, -50 to +50°Cdp
- High accuracy ±0.1°C fundamental reference
- User-configurable temperature and humidity profiling allows calibration cycles to be run without constant supervision
- Very fast changes between generated relative humidity points
Technical Specification
| General | |
| Enclosure | 19” Rack System, H= 2.1m |
| Power Supply | 100 to 115 V or 220 to 240 V 50/60Hz |
| Pressure Swing Dryer | |
| Gas Output | Flow: 7 Nl/min Pressure: 0.68 barg Moisture Content: <1ppmV (<-75°Cdp) |
| Required Gas Supply | Flow: 10 Nl/min Pressure: 5 to 7 barg Moisture Content: Oil and liquid water free |
| Type | Twin column desiccant, pressure swing |
| Desiccant | 4 Ångström Molecular sieve bead (4-8 mesh) |
| Timer | Motorized cam |
| Operating Temperature | +5 to +35°C |
| Generator | |
| Generation Range | Humidity: 1 to 95% RH (-50 to +50°Cdp) Temperature: Dependant on temperature chamber |
| Generated Gas Output | Air 2 Nl/min @ 0.5 barg via heat traced line |
| Dual Stage MFC Mixing | Dual mass flow controllers |
| Power Consumption | 550 V A maximum |
| Dual Stage MFC Mixing | Dual mass flow controllers |
| Operating Temperature | +5 to +40°C; 10 to 90% RH |
| Enclosure | 19″ Rack System, H= 2.1m |
| Control System | Closed loop feedback |
| Reference Instrument | |
| Performance | |
| Measurement Technology | Chilled Mirror |
| Measurement Range | -80 to +85°Cdp |
| Measurement Accuracy | ±0.1°Cdp ±0.1° temperature |
| Repeatability | Better than ±0.1°C |
| Resolution | 0.01 (0.1 for % RH) |
| Dew-Point Sensor | |
| Mirror | Gold plated copper |
| Temperature Measurement | 4 wire Pt100, 1/10 DIN class B |
| Sample Flow Rate | 0.1 to 0.7 Nl/min (recommended) |
| Integrated Flowmeter | 0 to 1 Nl/min |
| Sensor Pressure | Atmospheric |
| Configuration | Remote |
| Remote PRT | |
| Temperature Measurement | 4 wire Pt100, 1/10 DIN class B |
| Monitor | |
| Resolution | 0.01°C |
| Dual Optics Detection | Wide band red LED with dual photo sensors, all system insulated |
| Outputs | Analog: 2 channels, 10 mV/°Cdp, 4-20 mA Digital: RS232 Logic: Data hold, ABC Logic status, optics alarm |
| Operting Temperature | 0 to + 40°C |